Common Practices and Procedures Not Found in
the Text
Experiments are reported an informal style, yet
the information is complete. Organization of the data provides
not only a clear but also a logical report. In the Example Tables
below a format is given that is used in several of the
experiments in this and future courses.
Example Reaction Table
This is a reaction with stoichiometry more
complex than you will encounter in this course, but the example
is instructive. The ultimate goal is to determine the theoretical
yield, given the amounts of reactants. Pyridine (C5H5N)
is converted with CrO3 and HCl to form an intermediate
compound, pyridinium chlorochromate, that is then used to oxidize
1-decanol (C9H19CH2 OH) to
decanal (C9H19CHO). The overall reaction
may be written without needing to write the formula for
pyridinium chlorochromate: C5H5N + CrO3
+ HCl + C9H19CH2 OH --> C9H19CHO
+ CrCl3 + H2O + C5H5NHCl
and when balanced: 2C5H5N + 2CrO3
+ 8HCl + 3C9H19CH2 OH -->3C9H19CHO
+ 2CrCl3 + 6H2O + 2C5H5NHCl.
In this example the procedure calls for 1.0 mL of pyridine, 4g of
chromium trioxide, 6 mL of 6M HCl and 1.0 mL of decanol.
| 2 C5H5N | 2 CrO3 | 8 HCl | 3 C9H19CH2OH | 3 C9H19CHO | |
| vol. used | 1.0 mL | 6 mL (aq) | 1.0 mL | ||
| density | 0.978 g/mL | 0.829 g/mL | |||
| mass | 1.0 mL x 0.978 g/mL =0.978 g = 978mg |
4g = 4000mg | 1.0 mL x 0.829 g/mL = 0.829g = 829 mg |
5.2 mmole x156 mg/mmole = 811mg = theoret. yield |
|
| MW | 79 mg/mmole | 100 mg/mmole | 158 mg/mmole | 156 mg/mmole | |
| Molarity | 6 mmole/mL | ||||
| mmoles | 978 mg ÷79 mg/mmole =12.0 mmole |
4000 mg ÷100 mg/mmole = 4.0 mmole |
6 mL x6 mmole/mL = 36 mmole |
829 mg ÷158 mg/mmole = 5.2 mmole |
5.2 mmole |
| calc. lim. reagent |
12 ÷ 2 = 6 | 4.0 ÷ 2 = 2 | 36 ÷ 8 = 4.5 | 5.2 ÷3 = 1.7 limiting reagent |
Densities, molecular weights are obtained from
the Handbook. Note that many of the blanks are not filled in
because the data is not necessary. Also the mmoles of HCl are
calculated from a VxM = moles calculation since HCl is in aqueous
solution. The "calculate lim. reagent" line is usually
not necessary, but since the ratio of coefficients in the
balanced equation is not 1:1:1 etc., another line calculation
must be made; when the millimoles of decanol is divided by its
coefficient the result is lower than the result for any other
reactant, therefore decanol is the limiting reagent.
Note also how the calculations can be made
right in the table. The logic is followed with arrows so that the
limiting reagent and the theoretical yield, both labeled, are
easily found.
Example Table of Chemical Tests.
This table is appropriate for the
identification of unknowns. Observe how the IR information is
included into the line of reasoning and suggests the chemical
tests. Even though the IR clearly indicates an alcohol,
preliminary alkene and alkyl halide tests are run anyway, since
negative tests provide confirmation. But not all chemical tests
need be run, only the ones suggested by the IR and preliminary
tests. The alkyl halide (Beilstein) test, like the Lucas test,
was not decisive the first time, and tests were run to check the
initial findings. The road to determining the functional group is
often laborious. Expect conflicts and to repeat tests.
| Test | Result | Inference | Comments |
| IR | rounded peak - large 3350 cm-1 |
alcohol | |
| IR | no peaks 3100-3000 cm-1 | not an alkene, aromatic | |
| IR | peaks 300-2800 cm-1 | sp3 C-H | alkyl |
| IR | no peak 1680-1750 cm-1 | no C=O | see McM Table 12.1 |
| IR | rounded peaks 1060 cm-1 | primary alc | see Mayo, Table 6.12 |
| Boiling Point | 151-155 deg Celsius | ||
| Beilstein | green flame | alkyl halide | alcohol which also has halogen ? |
| Br2/CCl4 | red persists | not an alkene | confirms IR |
| BP repeat | 158-160 deg Celsius | if alcohol - about 5-6 carbons | first time was less clear - this time I took my time |
| Lucas | no cloudiness -did not dissolve | more than 6 carbons | |
| (Beilstein on known pentanol) | green flame but very brief | false positive | recheck unknown |
| Beilstein repeat on unknown | very brief green | probably not alkyl halide | green not persistent |
| (Beilstein on known chlorobenzene) |
much stronger and longer green | real positive | unknown is definitely not like chlorobenzene that really has halogen |
| Chromic Acid | opaque blue-green suspension | primary or secondary alcohol | |
| ceric nitrate | yellow to red | alcohol of less than 10 carbons | From Lucas, BP and this, unknown is an alcohol having 5-10 carbon atoms |
| Redo Lucas using fresh reagent |
cloudiness in 3 min | secondary alcohol | other students noticed that the Lucas was not working - made a fresh solution. Therefore unknown can be 6 or less carbons? |
| Repeat Lucas using fresh reagent | cloudiness in 2.5 min | secondary alcohol | IR indicates primary. I cleaned the test tubes and pipettes. |
| Iodoform | (didn't have time) | not that important - only certain secondary alcohols will be positive |
Example Table of Candidates.
This table is made to organize the selection of
a single compound from a list that best fits the data. In this
example the progress is followed from a Table of Tests that
determined that the unknown was an alcohol having about 6 carbons
and was either primary (from the IR - but note that the 1060 cm-1
is a little high) or secondary (from the Lucas). So all
candidates alcohols having boiling points 5 degrees below and 15
degrees above 159 are listed.
| Candidate | Synonym | Structure | 1o 2o 3o | bp (oC) | Does This Fit My Spectrum ? |
| (My unknown) | ~164 | ||||
| n-hexyl alcohol | 1-hexanol | CH3CH2CH2CH2CH2CH2OH | 1o | 157.5 | peak @ 1050 cm-1too wide |
| 2-hepanol | CH3CH2CH2CH2CH2CH(OH)CH3 | 2o | 158 | has 1100 cm-1 peak -no | |
| cyclohexanol | [supply drawing] | 161.5 | best fit - small extra peak at 930 cm-1 | ||
| 2-methyl-1-hexanol | CH3CH2CH2CH2CH(CH3 )CH2OH | 1o | 164 | has large 1130 cm-1 - no | |
| 2-methylcyclohexanol | [supply drawing] | 2o | 167.4 | 1350 cm-1peak
too small too many small peaks bet 1050-1100 cm-1 |
|
| n-heptyl alcohol | 1-heptanol | CH3CH2CH2CH2CH2CH2CH2OH | 1o | 176.8 | 1380 cm-1 peak
too small 1050 cm-1 too fat |
In the following example the student was to
identify a compound that was either an alkane or an alkene. In
this experiment NMR data are available. Most of the chemical
tests and the preliminary spectral data in this example indicated
that it was an alkene. The first line indicates the student's
data. So all candidates having boiling points 5 degrees below and
15 degrees above 165 are listed, then to make sure, the boiling
point window was extended past these limits.
| Candidate | Structure | bp oC | Does This Candidate Fit My IR and NMR Spectra? |
| (My unknown) | (alkene) | 165 | |
| Ketone Candidates | |||
| 6-methyl-3-heptanone | [supply drawing] | 160 | not good - big peak at 1370 cm-1 |
| 2-methyl cyclohexanone | [supply drawing] | 163 | best - peak at 980 cm-1 a little large and split |
| 3-methyl cyclohexanone | [supply drawing] | 170 | not good - large pks bet 1200 and 1400 cm-1 |
| methyl cyclohexyl ketone | [supply drawing] | 180 | fair - peak at 1333 should be at 1300 cm-1 |
| 2-octanone | [supply drawing] | 173 | not good - large wide peak at 1360 cm-1 |
| cyclohexanone | [supply drawing] | 156 | close - extra peaks at 1430, 1330, 1140 missing pks at 990 and 960 cm-1 |
| 2-heptanone | [supply drawing] | 151 | not good - large peak at 1360 cm-1 |
| hexane-2,5-dione | [supply drawing] | 188 | way off - large rounded peaks at 1370 and 1175 cm-1 |
| Aldehyde Candidates | |||
| furfural | [supply drawing] | 162 | no - extra peak at 1725 cm-1 |
| heptanal | [supply drawing] | 155 | no - has extra peak at 1725 cm-1 |
| benzaldehyde | [supply drawing] | 179 | no! 1725 cm-1 |
| 5-methylfurfural | [supply drawing] | 187 | no! 1725 cm-1 |
All candidates must have simple drawings of
their structures (only some, above, could be printed in this
example.) Structures are found in the Handbook in an appendix
"Structural Formulas of Organic Compounds" to the list
"Physical Constants of Organic Compounds"
A number of techniques are introduced in the
first laboratory. Below are some more detailed instructions that
will be referred to in later laboratories.
Handling of Liquids. See Mayo, et al., p.40-45 Microscale amounts of liquids are not poured but instead transferred with a pipette. In your drawer are Pasteur pipettes and rubber bulbs - which is used for most work; also in the drawer are a graduated pipette (which fits into the green pi-pump) and a syringe.
As explained in the Check-In lab, Pasteur pipettes are held in the palm of the hand with the smallest finger while the rubber bulb is squeezed with the thumb and forefinger. This technique gives the best control of the pipette tip when the other hand is holding the vessels holding the liquids.
During any transfer with a pipette some liquid may leak out. To prevent loss of this liquid it is important to bring the mouths of the dispensing and receiving vessels as close as possible together.
The Pasteur pipette holds about 2 mL and is easily cleaned for further use. If more than a few milliliters of acetone are not enough to clean the pipette it is cheaper to throw it away and replace it from the stockroom.
The 2 mL graduated pipette/pi-pump combination
measures with more accuracy, but when you hold it with your thumb
on the roller it is clear that the pipette tip is much less
steady. The other hand may need to hold the tip, but this will
not allow you to hold the source or destination containers. In
addition, the graduated pipettes are not so cheap to replace.
Handling of Solids. Microscale
quantities of solids are weighed on ordinary pieces of paper; but
these can be folded to channel the solids as they are then poured
into the test tube, flask, etc. containers. Because small amounts
are used it is advantageous to precut the paper so the solid runs
along a fold directly to the bottom of the vessel without
touching its neck. For emptying vessels use of the microspatula
requires practice and patience.
Boiling Point. The micro technique will
be described in a future procedure.
Recrystallization. This is a technique
for purification of solids, Generally, a solvent is selected
that, when boiling, will dissolve the solid. When the solution is
then cooled, the solvent will not be able to dissolve the solid
as well. As a consequence, purified solid will precipitate out of
solution as crystals. These crystals are recovered by filtering
them from the solvent; the separated solvent carries the
impurities with it. Specific instructions will be given in
procedures.
Routine use of the Gas Chromatograph. Text p. 66-70. Instrument is a Gow-Mac with a Carbowax 20 M (polar) column and a DC-200 (non - polar) column. Setting up an instrument such as the gas chromatograph is a skill learned in an advanced course, but the parameters of the analysis must be measured and written right on the chart: The carrier gas, helium, with a back pressure at about 18 pounds per square inch read at the gauge and a flow about 20-25 ml/min read at the flowmeter. The column used, the oven temperature, the attenuation, the nature of the sample and the amount injected, the attenuation (i.e. the sensitivity of the detector) and the chart speed. (about 1 inch per minute; this should be verified with a ruler and a clock.). The instrument will be set up for your work; make sure that the detector power is on only when helium is flowing.
Prior to your injection, adjust the pen, with the chromatograph zero knob, to the right side of the main chart (the small chart on the right side of the recorder is not used.)
To inject, take the automatic syringe and clean it several times by filling and emptying it several times with your sample; discharge the syringe into the air. Adjust the barrel of the syringe to the microliter (µL = 0.001 L) capacity desired. For normal runs 3 µL is typical. Take up the sample into the syringe, and inject it into the injector port corresponding to the desired column. Make a mark with a pen on the chart at the point of injection. While you await peaks, make sure you have documented the parameters underlined above on the chart.
The first injection may cause the chromatogram to be too big or too small. To make the peaks larger, increase the sensitivity by lowering the attenuation. If your peaks of interest are too large, turn up the attenuator. Often large peaks which are not of interest, are allowed to pin the pen so that smaller peaks can be seen and measured. This will be the case with the free radical chlorination experiment.
If another student is about to inject, permit him/her to inject and wait until your chromatogram has passed the sprocket. Now tear off your chart, take it to your bench and staple it to an empty page in your notebook. Identify as many of the peaks as possible by marking them. The instructions for measuring the retention time and integrating each peak are given in the "check-in" lab.
Calculations of the retention times for the peaks of interest must be put on the chromatogram or on the same notebook page that the chromatogram is attached to. Since the chart speed is 1 inch per minute, it is easiest to measure (in millimeters) the horizontal distance between the injection point and the point when the peak has reached its maximum. Multiply this distance by (minute/25.4 mm) to obtain the retention time. Repeat this for all peaks needing to be measured.
Calculations for the areas for each peak to be measured - also written on the chromatogram - are done by the height times width-at-half-height method: 1) draw a baseline that would have resulted if the peak had not appeared; 2) from the baseline measure and record the height, the vertical distance in millimeters from the baseline to the top of the peak; 3) mark the midpoint of this line; 4) measure the width of the peak, through the midpoint mark (i.e. the width-at-half-height). Mulitply the height by the width-at-half-height and record the result. The area of each measured peak is then calculated as a percent of the total measured peaks.
Taking an Infrared (IR) Spectrum . McM sections 12.4-12.8. (the Mayo, et al. treatment in Ch. 6 need only be scanned since it is overly detailed for an introduction). The procedure the preparation and measurement of a liquid (neat) sample is found in the "check-in " lab file.
Solids melting below 80 oC are run as "melts": a clean salt plate is loaded with 3-5 mg of the solid and the plate is carefully warmed on a clean, dry surface (aluminum foil on a warm hot plate) until the solid is melted. The sample is mounted and run as usual.
Solids melting above 80 oC are run as KBr pellets: 1) obtain a small agate mortar and pestle from the stockroom, and grind 3-6 mg of sample and about 50 mg of KBr until the solid is caked and glassy. 2) With a spatula, transfer the mixture to a clean, minipress (Mayo, et al., Figure 6.45); the instructor will then press the sample into a window that will be mounted in the spectrophotometer and the sample is run as usual.
Before the chart is removed from the instrument the spectrum is calibrated to correct any shift of the spectrum due to an incorrect alignment of the spectrum chart on its tray. The sample is replaced by a film of polystyrene (in a card) in the sample beam and run between 3200 and 2900 cm-1. The scan for the polystyrene should be started with the pen up at 3200. Using the spectrum of polystyrene taped to the instrument as a reference, observe the strong absorption beginning at about 3100 and continuing with the pen bottoming out briefly at 3026 then seriously at about 2924. Prepare to put the pen down, while scanning, just long enough to catch the peak at 2850, the calibration peak. Then raise the pen just after the peak. If this peak is actually marked on the chart at 2850, no adjustment of your spectrum is necessary. If, however, the calibration peak falls above 2900 or below 2800 on the chart, you must adjust the peaks up or down in the same direction (and amount) that the calibration peak would need adjusting to record 2850. Because of a scale change at 2000 cm-1 any correction below 2000 is half the number of cm-1 than the correction at 2850. Fully document your calibration with notations on the spectrum.
Every spectrum must be identified by marking
the Name, Sample (for example "aldehyde/ketone unknown #
034), Date, Class and Method (and when appropriate the Extract
and Solution.)
Interpreting an IR Spectrum. In the
"check-in" lab, identification of the unknown substance
only requires comparison of the spectrum obtained with 25 spectra
in chapter 6 of Mayo, et al.. All future spectra must be
interpreted and annotated even if a match happens to be found. In
McM, section 12.4-12.5 you will find a quick introduction to IR
theory; vibrations of the bonds within molecules give rise to the
various peaks in IR spectra, and therefore analysis of the
spectra will give information on the bond sequences or functional
groups within the compound being scanned.. Mayo, et al., p. 231,
gives a good strategy on organizing this information.
1. Measure and mark the wavenumber values
(corrected, if necessary, after calibration) for all significant
peaks between 4000 and 1350 cm-1. Remember that the
wavenumber values, not so much the intensities, of these peaks is
important. Also note the shapes (rounded, sharp, wide, narrow) of
the peaks.
2. If solvents or Nujol were used to mount the sample, they must be marked out. Spectra for these substances are found in the Aldrich IR Library
3. Interpretation of the spectrum itself is best done by first checking major features for fundamental functional groups. There are 3 regions to scan initially. These are best followed from McM Table 12.1and the figures in Mayo, et al., Chapter 6:
a) the O-H and N-H stretching region between 3600 and 3200 cm-1 An alcohol is very easily detected by a very large parabolic peak in this region (Fig 6.24), while an amine N-H peak is less rounded and dominating and for primary amines appears as two peaks (Fig 6.32 - see also Fig 6.35). The carboxylic acid O-H peak is entirely different; it is rounded and strong, but it is also very wide, extending from 3500 to 2500 cm-1 (Fig 6.29).
b) the C-H stretching region between 3100 and 2850 cm-1 is conveniently divided down the 3000 cm-1 line. Almost invariably the peaks between 3000 and 2850 are due to bonds between H and sp3 hybridized carbon atoms; this H-C stretch is present in every figure in Chapter 6 except Figs 6..10b (DCCl3) and 6.39 (chlorobenzene). Absorptions between 3100 and 3000 cm-1 arise from bonds between H and sp2 hybridized carbon, as in alkenes (Fig 6.15) and aromatic compounds (Fig. 6.39. Chlorobenzene is aromatic because it contains a benzene ring, a group of 6 carbons bound in a ring with alternating double and single bonds. Details of this group will be given in Chem 3020, but its spectral characteristics are easily recognized).
c) the carbonyl (C=O) stretch between 1690 and 1760 cm-1. This feature is in carboxylic acids, aldehydes, ketones, esters and amides.
d) other distinguishing peaks are in McM table
12.1. The functional groups listed are of less importance than a)
- c) above. Most of the functional groups listed has an example
in one of the Figures in Mayo, et al., Chapter 6.
4. The Handbook has "Infrared Correlation
Charts" which are helpful. In the Aldrich IR Library the
spectra are organized by functional group (alkenes, alcohols,
etc.); each section begins with a general discussion of the
characteristic peaks that are present in the functional group.
5. Remember to look first at the most
outstanding aspects of the spectrum, not get into minor details.
Do not try to interpret every peak in the spectrum.
6.Also remember that the absence of a peak is
often just as important as finding one, for it will narrow the
list of possibilities for your compound.
7. Now that the major peaks have been identified, further confirming and sub-classifying peaks can be located. The best strategy for this is to refer to the appropriate section in McM. Always check example spectra in McM and in Mayo, et al. Details of groups that may be encountered in this course:
a) alcohols - Mayo Table 6.12. Examination of the C-O stretching peak is useful for narrowing the alcohol to either a primary, secondary or tertiary alcohol (these figures often do not apply to cyclic alcohols).
b) alkenes - McM Section 12.7.
In general, once a functional group is
identified, specifics on its IR characteristics can be confirmed
and sub-classified by looking up the group's IR data in the
"spectroscopic analysis" section of the appropriate
chapter of McM.
8. A match can be sought from the Aldrich
Libraries to obtain a final identification. [Aldrich library of
Infrared Spectra and The Aldrich Library of FT-IR Spectra by C.J.
Pouchert, Aldrich Chemical Publishing Co. Milwaukee, WI]. For an
exact match there should be a peak for peak correspondence
between your spectrum and the library entry. Note, however, that
the intensities will differ if the amounts sampled are not
the same. Also the wavenumber values should match even though
scaling is different; the spectrum you obtain is linear in
wavenumbers (cm-1 values) while the Aldrich Library
spectra are linear in microns, so each major peak should be
measured separately. A handy conversion: wavenumbers =
10,000/microns.
9. Mark your spectrum liberally. Make notations
on each diagnostic peak (see McM Figure 12.11for example) and for
each peak that you used to arrive the structure of your unknown.
10. If two candidate compounds you found in the
Aldrich Library are both almost identical to your spectrum, make
a peak-for peak comparison. To do this mark off two 0.5-1 cm-wide
bands along the length of your spectrum. Label each band is
labeled with a candidate's name and insert small vertical lines
in the positions corresponding to that candidate's peaks (heavy
lines for strong peaks, etc.) All the information is now on your
spectrum and you should be able to determine the best fit for
your unknown.
References:
Text, frequently "Mayo" in the procedures: Mayo, D.W., Pike, R.M., Butcher, S.S. and Trumper, P.K. Microscale Techniques for the Organic Laboratory; Wiley: New York, 1991
Text for the Lecture, frequently "McM" in the procedures: McMurry, J. Organic Chemistry, 4th ed., Brooks/Cole Publishing Company, Pacific Grove, CA. 1996
The "Handbook" refers to any of the recent editions of: Weast, R.D. Handbook of Chemistry and Physics; The Chemical Rubber Co.: Cleveland, 1960-present.
The "Aldrich IR Library"
refers to any edition of: Pouchert, C.J. The Aldrich Library
of Infrared Spectra; Aldrich Chemical Co. Milwaukee,
1970-present.
Rev. December, 1998